6054/6055/7007 Deep-Well Baths

Fluke · 6054

6054/6055/7007 Deep-Well Baths

Request a quote

Category: Measuring instruments · Temperature calibration

Specifications by model

Characteristic700760546055
Range–5 °C to 110 °C50 °C to 300 °C200 °C to 550 °C
Temperature stability± 0.001 °C at 0 °C (water/ethanol) ± 0.001 °C at 25 °C (water) ± 0.003 °C at 100 °C (oil 5012)± 0.003 °C at 100 °C (oil 5012) ± 0.005 °C at 300 °C (oil 5017)± 0.003 °C at 200 °C (salt) ± 0.01 °C at 550 °C (salt)
Temperature uniformity± 0.004 °C at 0 °C (ethanol) ± 0.007 °C at 100 °C (oil 5012)± 0.007 °C at 100 °C (oil 5012) ± 0.015 °C at 300 °C (oil 5017)± 0.005 °C at 200 °C (salt) ± 0.010 °C at 550 °C (salt)
Temperature SettingDigital display with push-button data entry
Set-Point Resolution0.002 °C, high res. 0.00003 °C0.01 °C, high res. 0.00018 °C
Display Temperature Resolution0.01 °C
Digital Setting Accuracy± 1 °C
Digital Setting Repeatability± 0.005 °C± 0.01 °C
Heaters250 to 1000 W250 to 1000 W260 to 2080 W
Working Area178 mm dia. (7 in)196 mm dia. (7.7 in)107 mm dia. (4.2 in)
Depth24 in24 in24 in
Wetted Parts304 stainless steel
Power230 V ac (± 10 %), 50 or 60 Hz, 15 A (Specify frequency, contact Fluke Calibration if CE mark required.)230 V ac (± 10 %), 50/60 Hz, 15 A max230 V ac (± 10 %), 50/60 Hz, 15 A
Volume42 liters (11.2 gal)50 liters (13.2 gal)19.8 liters (5.2 gal, 43 kg [95 lb] of bath salt)
Size (D x W x H)470 x 775 x 194 mm to working surface (18.5 x 30.5 x 47 in), 1397 mm (55 in) to top of stir motor, 914 mm (36 in) to control panel572 x 762 x 1219 mm to working surface (22.5 x 30 x 48 in), 1422 mm (56 in) to top of stir motor box, 914 mm (36 in) to control panel572 x 775 x 1219 mm to working surface (22.5 x 30.5 x 48 in), 1524 mm (60 in) to top of stir motor box, 914 mm (36 in) to control panel
Distance from Line of Sight to Top of Fluid9.5 mm (3/8 in)15.9 mm (5/8 in)
AutomationUse the 1586A Super-DAQ to automate temperature sensor calibration. For IEEE-488, add the 2001-IEEE to the automation package.
Weight70.8 kg (156 lb)

Full description

Key features: Extra-deep wells suitable for thermometry work requiring additional tank depth and ultimate stability. Constant liquid levels maintained through a concentric-tube design. Special design tailored for the sighting of Liquid-In-Glass (LIG) thermometers. Well depths varying from 17 to 24 inches, eliminating stem conduction effects in probes requiring deep immersion.

In the same family