Fluke 700PCK Pressure Calibration Kit
Price on request
Details

Fluke · RPM4 mhhhac/mlllac
Category: Measuring instruments · Pressure calibration
| Characteristic | RPM4 BA100K (Single Channel) | RPM4 BA100K/BA100K (Dual Channel) |
|---|---|---|
| Range | 70 kPa to 110 kPa (10 psia to 16 psia) | |
| Power requirements | 85 V ac to 264 V ac, 50/60 Hz; 12 V dc, 1.2 A (battery) | |
| Weight | 5 kg (11 lb) | |
| Dimensions (H x W x D) | 10 cm x 22.7 cm x 24 cm (3.9 in. x 8.9 in. x 9.5 in.) | |
| Test port connections | 1/8 in NPT F | |
| Communications ports | RS232 (COM1, COM2), IEEE-488.2 Resolution To 1 ppm, user adjustable To 1 ppm, user | |
| Resolution | To 1 ppm, user adjustable | |
| Operating temperature | 15 °C to 35 °C (59 °F to 95 °F) | |
| Warm up time | 30 minute temperature stabilization recommended from cold power up | |
| Vibration | Meets MIL-T-28800D | |
| Predicted Stability1 | ± 0.005 % of reading | ± 0.0032 % of reading |
| Precision2 | ± 0.008 % of reading | ± 0.006 % of reading |
| Measurement Uncertainty3 | ± 0.01 % of reading | ± 0.008 % of reading |
| 1 Predicted Q-RPT measurement stability limit (k=2) over one year assuming regular use of AutoZero function. AutoZero is performed by the operator by comparison with a barometric reference in absol mode. Absolute mode predicted one year stability without AutoZ is: Single Channel ± (5 Pa + 0.005 % of reading), D Channel ± (4 Pa + 0.0032 % of reading). 2 Combined linearity, hysteresis, repeatability. 3 Maximum deviation of the Q-RPT indication from the true value of applied spressure including precision, predicted one year stability limit, temperature effect and calibration uncertainty, combined and expanded (k=2) following the ISO “Guide to the Expression of Uncertainty in Measurement.” |