Fluke · 5606-50-P
5606 Full Immersion PRT
Prix sur demande
Détails

Fluke · CAL-1595A
Catégorie : Instruments de mesure · Étalonnage de température
| Caractéristique | Valeur |
|---|---|
| 1595A General Specifications | |
| Measurement capability | 4-wire PRT, Thermistor, Resistance, Resistance Ratio |
| Input resistance range | 0 to 500 kΩ |
| Ratio range | 0 to 10 |
| Accepted external RS range | 1 Ω to 10 kΩ |
| Internal RS | 1 Ω, 10 Ω, 25 Ω, 100 Ω, and 10 kΩ |
| PRT conversion types | ITS-90, PT-100, CVD-ABC, CVD-ALPHA, Polynomial |
| Thermistor conversion types | R(T) Polynomial, T(R) Polynomial |
| Display units | ratio (RX / RS), K, °C, °F, Ω |
| Résolution d'affichage | 0.1 to 0.000001 |
| Sample period | (seconds)1 1, 2, 5, and 10 |
| Statistics | Average, Std Dev, SE of Mean, Max, Min, Difference, Peak-Peak, Delta, N |
| Front panel channels | Four PRT/Thermistor inputs (channels 2 & 4 can be configured for either RX or RS inputs) |
| Rear panel channels | Two dedicated reference resistor inputs (RS) |
| Input terminals | DWF Connector, Tellurium Copper |
| Data logging to internal memory | 80,000 individual time- and date-stamped readings (~6 MB) |
| Internal real-time clock | Oui |
| Computer communications | RS-232, USB, IEEE-488, Ethernet |
| Display type | Full VGA, LCD |
| User interface languages | English, French, Spanish, German, Russian, Chinese, Japanese |