Specifications by model

CharacteristicValue
ModelsNP-5500, NP-5500C, NP-5600, NP-5600C, NP-5800, NP-5800C
Light sourceLampe xénon flash longue durée (10⁹ éclairs)
Required volumeUltra-trace : 0,3–2 µL
DetectorCapteur CMOS linéaire 2048 éléments ou barrette CCD silicium 3864 éléments (selon modèle)
Detection range200–850 nm (200–800 nm selon modèle)
λ and absorbance accuracy±1 nm ; 0,002 Abs
Spectral resolution (bandwidth)2 nm (FWHM à Hg 546 nm)
Operating systemWindows 10 ou LINUX (selon modèle)
ScreenÉcran tactile 7'' / 8'' / 10,1'' capacitif (selon modèle)

Full description

Nano Prime Series Microvolume UV-Vis Spectrophotometer Empowering innovation and redefining the performance boundaries of micro-sample analysis The Nano Prime series Microvolume UV-Vis Spectrophotometer redefines the precision and efficiency in analytical workflows. Utilizing advanced dual-channel technology and intelligent detection systems, the NanoPrime series seamlessly adapts to diverse research, quality control, and compliance requirements—from academic laboratories to pharmaceutical manufacturing. Whether for nucleic acid/protein quantification, colorimetric analysis, kinetic studies, colloidal gold detection, or other diverse applications, the NanoPrime series delivers reliable, accurate results with a rapid workflow and professional data management. Nano Prime Series Microvolume UV-Vis Spectrophotometer Core Feature Extended Detection Range and High Precision: Dual-channel full-wavelength detection for microvolume and cuvette modes addresses broader absorption spectrum research needs. With 2 nm wavelength resolution and 0.002 Abs absorbance accuracy, capable of measuring 2–15000/2–30000/2–37500 ng/μL (dsDNA). it precisely measures concentrations across the entire range from ultra-trace levels to high-concentration samples. Intelligent Pollutant Identification and Result Correction: NP-5800 and NP-5800C Equipped with an exclusive and advanced intelligent sample detection system that automatically identifies common contaminants in samples (such as proteins, nucleic acids, phenols, guanidine salts, etc.) and provides contamination alerts, type identification, and original concentration result correction functions. Provides reports on the original concentration, corrected concentration, and contaminant type, fundamentally ensuring sample quality and data accuracy, and helping you obtain optimal experimental results. Comprehensive application module coverage: Nucleic Acid Analysis: dsDNA, ssDNA, RNA, Oligos, and gene chips Protein Quantification: Protein assay & purity analysis (260/280 ratio). BCA, Bradford, Lowry, and Pierce 660.