Secondary PRT with Calibration Options - 5608 and 5609

Fluke · 5608, 5609 PRTs

Secondary PRT with Calibration Options - 5608 and 5609

Request a quote

Category: Measuring instruments · Temperature calibration

Specifications by model

CharacteristicValue
Specifications
Temperature range5608: –200 °C to 500 °C 5609: –200 °C to 670 °C
Nominal resistance at 0.01 °C100 Ω ± 0.5 Ω
Temperature coefficient0.0039250 Ω/Ω/°C
Relative Accuracy [1]± 0.007 °C at –200 °C ± 0.008 °C at 0 °C ± 0.020 °C at 420 °C ± 0.027 °C at 660 °C
Short-term repeatability [2]± 0.007 °C at 0.010 °C ± 0.013 °C at max temp
Drift [3]± 0.01 °C at 0.010 °C ± 0.02 °C at max temp
Hysteresis± 0.01 °C maximum
Sensor length30 mm ± 5 mm (1.2 in ± 0.2 in)
Sensor location3 mm ± 1 mm from tip (0.1 in ± 0.1 in)
Sheath materialInconel™ 600
Minimum insulation resistance5608: 500 MΩ at 23 °C, 20 MΩ at 500 °C 5609: 500 MΩ at 23 °C, 10 MΩ at 670 °C
Transition junction temperature range [4]–50 ºC to 200 °C
Transition junction dimensions71 mm x 12.5 mm (2.8 in x 0.49 in)
Minimum immersion length [5] (5608: 80 mm (3.1 in) 5609: 100 mm (3.9 in)
Maximum immersion length305 mm (12 in)
Response time [5]5608: 9 seconds typical 5609: 12 seconds typical
Self heating (in 0 °C bath)5608: 75 mW/°C 5609: 50 mW/°C
Lead-wire cable typePTFE 24 AWG
Lead-wire length1.8 m (6 ft)
Lead-wire temperature range–50 °C to 250 °C
Calibration
[1] Includes short-term repeatability and 100 hr drift. Calibration will add additional uncertainties. [2] Three thermal cycles from min to max temp, includes hysteresis, 99.9 % confidence (k=3) [3] After 100 hours at max temp, 99.9 % confidence (k=3) [4] Temperatures outside this range will cause irreparable damage. For best performance, transition junction should not be too hot to touch. [5] Per ASTM E 644Calibration Uncertainty for accredited† calibration
TemperatureExpanded Uncertainty (k=2)
–197 °C0.025 °C
–80 °C0.025 °C
–38 °C0.025 °C
0.01 °C0.025 °C
156 °C0.030 °C
231 °C0.030 °C
419 °C0.035 °C
660 °C0.050 °C
Note: Calibration uncertainties depend on the uncertainties of the lab performing the calibration. Subsequent calibrations of this probe performed with different processes, at different facilities, or with changed uncertainty statements may state different uncertainties. †Lab code 200348-0

Full description

Key features: 5608: –200 °C to 500 °C (80 mm minimum immersion) 5609: –200 °C to 670 °C (100 mm minimum immersion) ± 0.01 short-term stability Calibration not included, accredited calibration optional, lab code Dry-well reference with 90 ° bend